Morgan Lens Instructions:
Instructions for using the Morgan Lens for continuous medication or lavage to the cornea and conjunctiva.
Download the Printer-Friendly PDF of these instructions (2 MB) »
.jpg)
Step 2: Attach Morgan Lens Delivery Set, I.V., set-up or syringe using solution and rate of choice; START FLOW. *

Step 1: INSERTION: Instill topical ocular anesthetic, if available.

Step 4: Release the lower lid over Morgan Lens and adjust flow. Tape tubing to patient's forehead to prevent accidental lens removal. Absorb outflow with the Medi-Duct. DO NOT RUN DRY.

Step 3: Have patient look down, insert Morgan Lens under upper lid. Have patient look up, retract lower lid, drop lens in place.


Step 5: REMOVAL: CONTINUE FLOW, have patient look up, retract lower lid - hold position.
Morgan Lens Recommended Uses:
| Ocular injury due to acid burns or solvents, gasoline, detergents, etc. | Solution: | Mode: |
| Lactated Ringer's** I.V. Solution | Morgan Lens Delivery Set or I.V. set-up | |
| Rate: | Frequency: | |
| 500 ml rapid/free flow. Reassess and continue at slower rate. | Once. Repeat as necessary. | |
| Alkali burns | Solution: | Mode: |
| Lactated Ringer's** I.V. Solution | Morgan Lens Delivery Set or I.V. set-up | |
| Rate: | Frequency: | |
| 2000 ml rapid/free flow. Reassess. Continue at 50 ml/hour or 15 drops/minute. | Continuous until pH of cul-de-sac is returned to neutrality. | |
| Non-embedded foreign bodies |
Solution: | Mode: |
| Lactated Ringer's** I.V. Solution | Morgan Lens Delivery Set or I.V. set-up | |
| Rate: | Frequency: | |
| 500 ml rapid/free flow. Reassess and continue at slower rate. | Once. Repeat as necessary. | |
| Foreign body sensation with no visible foreign body | Solution: | Mode: |
| 20 cc sterile solution | 20 cc syringe | |
| Rate: | Frequency: | |
| Slowly without force. | Once. Repeat once if necessary. | |
| Routine pre-operative | Solution: | Mode: |
| 10 cc of preferred ocular antiseptic | 10 cc syringe | |
| Rate: | Frequency: | |
| Slowly without force. | Once. | |
| Eyelid surgery | Solution: | Mode: |
| Lactated Ringer's** I.V. Solution | Morgan Lens Delivery Set or I.V. set-up | |
| Rate: | Frequency: | |
| 4 drops/minute. | During entire procedure. | |
| Severe infection | Solution: | Mode: |
| Lactated Ringer's** I.V. Solution with suitable antibiotic and steroid*** | Morgan Lens Delivery Set or I.V. set-up | |
| Rate: | Frequency: | |
| 50 ml/hour or 15 drops/minute. |
Continuous for 70 hours, then 10-hour intervals until marked improvement. | |
*Allows Lens to "float" over cornea and sclera.
**Recommendation based on pH: Tears approxiately 7.1, Normal Saline 4.5 to 7.0, Lactated Ringers 6.0 to 7.5.
***Use only when indicated.
Copyright © 2007 The Morgan
Lens by MorTan Inc. All Rights Reserved.
To provide you with the highest quality products,
MorTan has obtained the following quality certifications: ISO 9001, EN
46001, ISO 13485, CMDCAS. In addition, all of MorTan's products carry the CE
mark and are manufactured using the US Food and Drug Administration's current
Good Manufacturing Practices.

![]()